ST2826 Operation manual
Ver1.0
◇
12
Lp-D
Lp-Q
Lp-G
Lp-Rp
More 3/7
4)
Select and press a softkey to set measure function. If the softkey you require is not
displayed on the page, press More 3/7 to display the following softkeys.
Ls-D
Ls-Q
Ls-Rs
More 4/7
5)
Select and press a softkey to set measure function. If the softkey you require is not
displayed on the page, press More 4/7 to display the following softkeys.
R-X
Z-Өº
Z-Ө r
Rs-Q
More 5/7
6)
Select and press a softkey to set measure function. If the softkey you require is not
displayed on the page, press More 5/7 to display the following softkeys.
G-B
Y-Өº
Y-Ө r
Rp-Q
More 6/7
7)
Select and press a softkey to set measure function. If the softkey you require is not
displayed on the page, press More 6/7 to display the following softkeys.
DCR
More 7/7
8)
Select and press a softkey to set measure function. Press More 7/7, the softkey of step 1
will be displayed. If you forget tp select the measure function, just reset the function from
step 1to 8.
3.1.2Range
The range can be selected according to the impedance value of LCR component.
ST2826 has 10 AC ranges
:
1Ω, 10Ω, 30Ω, 100Ω, 300Ω, 1kΩ, 3kΩ, 10kΩ
、
100kΩ and 1MΩ.
ST2826 has 10 DCR ranges
:
1Ω, 10Ω, 30Ω,100Ω, 300Ω, 1kΩ, 3kΩ, 10kΩ,
、
100kΩ and 1
MΩ.