ST2826 Operation manual
Ver1.0
◇
22
Test frequency
(
Frequency
)
Test level
(
Level
)
DC bias
(
Bias
)
Test speed
(
Speed
)
Trigger mode
(
Trigger
)
Auto level control
(
Constant level
)
Bias current isolation ON/OFF (
ISO
)
Average time
(
Average
)
Voltage level monitor ON/OFF (
V
m
)
Current level monitor ON/OFF (
I
m
)
Delay time
(
Delay
)
Output resistance
(
Internal resistance
)
File management
(
File
)
Auxdiary tools
(
Tools
)
Deviation test mode A
(
Deviation
A)
Deviation test mode
B(
Deviation
B)
Deviation test reference
A(
Reference
A)
Deviation test reference
B(
Reference
B)
Some setting zones on
<
Measure setting
>
page is the same as that on
<
LCRZ
>
page
,
listed as below, which is introduced before. Other setting zones on
<
Measure setting
>
page
will be introduced briefly below.
Test function
(
Function
)
Test range
(
Range
)
Test frequency
(
Frequency
)
Test level
(
Level
)
DC bias
(
Bias
)
Test speed
(
Speed
)
3.4.7
Trigger mode
In ST2826 there are 4 types of trigger mode
:
INT(internal trigger)
,
MAN(manual trigger)
,
EXT(external trigger) and BUS(bus trigger).
When the trigger mode is set as INT mode
,
ST2826 can test constantly.
When the trigger mode is set as MAN mode, press
[TRIGGER]
once
,
ST2826 performs a
test.
When the trigger mode is set as EXT mode, once HANDLER receives a trigger signal of
positive pulse
,
ST2826 performs a test.
When the trigger mode is set as BUS mode
,
once IEEE488 receives a ‖
TRIGGER
‖
,
ST2826
Summary of Contents for ST2826
Page 2: ......
Page 10: ......
Page 51: ...ST2826 Operation manual Ver1 0 41 Figure 3 15 ...
Page 61: ...ST2826 Operation manual Ver1 0 51 ...
Page 144: ...ST2826 Operation manual Ver1 0 134 Figure 9 8 the input circuit for control signal ...