ST2826 Operation manual
Ver1.0
◇
53
o)
Press OFF, close the dot frequency clear function for frequency 2.
p)
Move the cursor to FREQ 3 field. ON, OFF, MEAS OPEN, MEAS SHORT and
MEAS LOAD will be displayed in the softkey area.
q)
Press OFF, close the dot frequency clear function for frequency 3.
5.2.1 Dot frequency clear (for single frequency)
Dot frequency clear for single frequency is recommended.
Suppose that the frequency in you test is 5.5KHz.
a)
Press [LCRZ], and press MORE1/3
.
MEAS SETUP, CORRECTION, LIMIT
TABLE, LIST SETUP, MORE2/3 will be displayed in the softkey area.
b)
Press the softkey CORRECTION. The CORRECTION PAGE will be displayed.
c)
Move the cursor to OPEN field. ON, OFF and MEAS OPEN will be displayed in the
softkey area.
d)
Press ON, enable the open correction function.
e)
Move the cursor to SHORT field. ON, OFF and MEAS SHORT will be displayed in
the softkey area.
f)
Press ON, enable the short correction function.
g)
Move the cursor to the LOAD field. ON and OFF will be displayed in the softkey
area.
h)
Press OFF, close the load correction function.
i)
Move the cursor to FREQ 1 field. ON, OFF, MEAS OPEN, MEAS SHORT and
MEAS LOAD will be displayed in the softkey area.
j)
Press ON, enable the dot frequency clear function for frequency 1.
k)
Press key [5],[.][5], +5.5 will be displayed in the bottom line of the screen, and the
usable units (Hz, kHz, and MHz) will be displayed in the softkey area. Press the
softkey kHz, the frequency 1 will be changed to 5.5000 kHz which is the same to the
test frequency.
l)
Keep the test fixture open, press the softkey MEAS OPEN to perform open
correction.
Summary of Contents for ST2826
Page 2: ......
Page 10: ......
Page 51: ...ST2826 Operation manual Ver1 0 41 Figure 3 15 ...
Page 61: ...ST2826 Operation manual Ver1 0 51 ...
Page 144: ...ST2826 Operation manual Ver1 0 134 Figure 9 8 the input circuit for control signal ...