ST2826 Operation manual
Ver1.0
◇
60
(d)Move the cursor to ALARM MODE field. The current field displays OUT
4) Connect the test fixture (ST26005) to ST2826‘s test terminals.
5) Perform the clear. (To avoid the disturbance of the stray capacitance, you should do
OPEN/SHORT CORRECTION by referring to Dot frequency clear in this chapter.)
6) Insert the capacitance to test fixture.
7) Perform the following measurement.
Press [LCRZ], and press LIST SWEEP.ST2826 will display LIST SWEEP DISP
page. The continually sweep test will be performed. The test and compared result will be
displayed. The beep will sound when the compared result is H (greater than the high) and
L(less than the low). Figure 5-4 shows the LIST SWEEP DISP page.
8) If the test result is error evidently, please perform the following steps:
(a)Check whether the capacitance is connected to the test fixture reliably
(b)Check whether the test fixture is connected to the test terminal of the instrument
reliably.
(c)Do open correction and short correction again.
Note: when you use frequency sweep open/short correction, the dot frequency correction
function should be OFF. Please refer to 5.1 in this chapter.
Figure 5-4 the LIST SWEEP DISP page
5.7Operation example for load correction
Procedures:
Suppose the test conditions are below:
frequency:100kHz
standard Cp:11nf
standard D:0.0005
(a)Press [LCRZ], and then press MORE 1/3. MEAS SETUP, CORRECTION, LIMIT TABLE ,
LIST SETUP and MORE 2/3 will be displayed in the softkey area.
Summary of Contents for ST2826
Page 2: ......
Page 10: ......
Page 51: ...ST2826 Operation manual Ver1 0 41 Figure 3 15 ...
Page 61: ...ST2826 Operation manual Ver1 0 51 ...
Page 144: ...ST2826 Operation manual Ver1 0 134 Figure 9 8 the input circuit for control signal ...