ST2826 Operation manual
Ver1.0
◇
75
The experiment instruments are shown in table 6-10
NO
The instrument name
requirement
1
Standard capacitance
100pF
0.02
%
(the loss D has been
known)
1000pF
10000pF
10nF
0.1uF
1uF
2
AC
standard resistor
10Ω
0.02
%
100Ω
1kΩ
10kΩ
100kΩ
3
DC
standard resistor
0.1Ω
0.02
%
1Ω
10Ω
100Ω
1kΩ
10kΩ
100kΩ
4
Standard capacitance
100μH
0.02%
1mH
10mH
100mH
5
Frequency meter
(
0
~
1000
)
MHz
6
digital multimeter
0.5
%
7
Insulation
resistance meter
500V 10 levels
8
withstanding voltage and leakage meter
0.25kW
(
0
~
500
)
V
6.6.3Check the Function
The keys, the display and the terminals can work normally. The other functions are right.
6.6.4The test signal level
When the test signal level is tested, the digital multimeter is used. Adjust the range of the