8.1.8BIAS subsystem command set
BIAS
subsystem command set is used to set the internal bias voltage.
Command hierarchical chart:
BIAS
:STATe ON (1)
OFF (0)
:VOLTage
<value>
MIN
MAX
(
仅
GPIB)
:CURRent
<value>
MIN
MAX
:STATe
is used to set the bias ON/OFF. Char ‗?‘ can be used to query the state of the bias.
Command syntax
:
ON
BIAS:STATe
OFF
1
0
char ‗1‘( 49) is equivalent to ON
char ‗0‘( 48) is equivalent to OFF.
e.g.
:
WrtCmd(―BIAS:STATe 0‖); set the DC bias close.
query syntax
:
BIAS:STATe
?
return
:
<NR1><NL^END>
:VOLTage
is used to set the bias voltage. Char ‗?‘ is used to query the current bias voltage.
Command syntax
:
<value>
BIAS:VOLTage
MIN
MAX
<value>: NR1,NR2 or NR3 format
MIN : set the bias voltage to 0V
MAX: set the bias voltage to 2V