29
Instruction:
Label
Instruction
Definition
PASS HOLD: 0.2S~99.9S
Pass judge hold time.
KEY
Press STOP key to quit holding.
STEP HOLD: 0.2S~99.9S
Waiting time.
KEY
Press ‘START’ to test the next item.
AUTO RANG: ON, OFF
Automatically switch the range 0.6s
before test end.
GR CONT:
OFF
No GND contact test in low terminal.
KEY
Press ‘START’ to do contact test.
0.2S~99.9S
Contact time setup.
BEEP VOL:
OFF, LOW, HIGH Alarm volume.
CONTRAST: 01~10
LCD contrast setup.
PASSWORD: ON, OFF, SET
Key lock setup. The default password
is 90000000.
GFI:
ON, OFF
Discharge wall function (see 4.4.7).
SYSTEM 2 Interface
SYSTEM2
AFTR FAIL:
STOP
STRT DLY1
OFF
LOCK
RAMP JUDG:
OFF
OFFSET
OFF
OFST
DC50 AGC:
OFF
DISP MODE
P/F
RMT
Part NO:
00000000
STRT DLY2
OFF
ERR
SAVE
F1
F2
F3
F4
F5
4.2.3.2 SYSTEM2
Instruction:
Label
Instruction
Definition
AFTR FAIL: STOP
Use ‘STOP’ to quit in fail result.
CONTINUE
Continue the next step in fail result.
RESTART
Retest in fail result.
RAMP JUDG: ON, OFF
Low limit judge in rising withstanding
voltage.
DC50 AGC: ON, OFF
High voltage hardware feedback in DC50V-
500V.
PART NO:
8 charcters
Number setup of the test programme.
STRT DLY1
OFF~0.1~99.9S Set the first test delay time.
OFFSET
OFF~ON
Activate the deduction of offset from the
test data.
GET
Obtain the current test data of the current
test condition and set it as offset.
DISP MODE
P/F
Large character display PASS and FAIL.
DATA
Small character display PASS and FAIL. No
influence on test data display.
STRT DLY2
OFF~0.1~99.9S Set the second test delay time.