17
Operation cont.
Fuse Replacement
1. Lift up the tilt stand.
2. Loosen the Phillips
screws on the battery/fuse
cover.
3. Remove the battery/fuse
cover.
4. Gently remove fuse and
install new fuse into the
holder.
5. Always use a UL recognized
fuse of the proper size and
value: 10A/600V (5 x 20mm)
fast blow.
6. Install the battery cover and
tighten the screws
To avoid electric shock, do not operate meter until
the battery/fuse cover is securely fastened to the meter.
WARNING:
To avoid electric shock, remove the test leads from
the meter before removing the battery/fuse cover.
WARNINGS:
Fuse
When replacing the battery or fuses, be sure to secure
the battery compartment door firmly to maintain the waterproof and dust
proof integrity of the meter. Loose or overtightened screws, or an improperly
seated o-ring may compromise the meter's water and dust ingress protection.
WARNING:
18
Specifications
4.000V
40.00V
400.0V
600V
All AC voltage ranges are specified 5% to 100% of range
Input Protection:
600V AC RMS or 600V DC
Input Impedance
7.8M
Ω
AC Response:
50 to 60Hz
1mV
10mV
0.1V
1V
±(1.5% +10 digits)
AC Voltage
Function
Range
Resolution
Accuracy
±
(
% of r digits)
400.0mV
4.000V
40.00V
400.0V
600V
Input Protection:
600V RMS or 600V DC
Input Impedance:
7.8M
Ω
0.1mV
1mV
10mV
0.1V
1V
±(0.8% + 8 digits)
±(1.2% + 2 digits)
±(1.2% + 5 digits)
DC Voltage
Function
Range
Resolution
Accuracy
±
(
% of r digits)
Accuracy is given at 65°F to 83°F (18°C to 28°C), less than 70% relative humidity
4.000V
40.00V
400.0V
All AC voltage ranges are specified from 5% of range to 100% of range
Input Protection:
300V AC RMS or 300V DC
Input Impedance:
approx.3k
Ω
AC Response:
50 to 60Hz
1mV
10mV
0.1V
±(1.5% +10 digits)
Low Z
AC Voltage
Function
Range
Resolution
Accuracy
±
(
% of r digits)