Rev 002, 04/11
5
© SP Scientific 2011
Chapter
3
Operation
Before beginning operation of the Lyo-SEA Sample Extractor, ensure that the front
plate is closed and that the inert gas/vacuum break valve is closed.
Operation
To operate the Lyo-SEA Sample Extractor:
Activate the auxiliary vacuum pump to depressurize the antechamber.
1.
After the gauge stabilizes at the maximum position, allow the vacuum pump to
2.
operate for an additional 15 minutes.
Open the antechamber isolation door. With locking handle in the up position
3.
(unlocked), slide the antechamber door open.
When the antechamber isolation door is fully open, you may lower the handle
4.
to lock the isolation door in position.
Antechamber Isolation Door
Raise handle to
unlock door.