Operating Your Lyophilizer
FTS Systems LyoStar™ II
22
Rev 012, 04/11
© SP Scientific 2011
To Change the Recipe While the System Is Executing the Program:
To advance to the next step, return to the automatic screen where the recipe
•
is displayed. Clicking on the green box of the step number that is currently
being executed will immediately force the program into the next step.
To modify the hold time of a current step, change the hold time to however
•
much longer you want to hold.
Examples:
You are in Step 1 below and the system has been holding for 10 minutes. You
decide you want to hold for only 5 minutes more. Change the hold time to 5.
You decide you want to hold for a total of 50 minutes and not 30. It has
already been 10 minutes. Change the hold time to 40 and the system will hold
for 40 more minutes for a total of 50 minutes.
Shelf temperature and vacuum setpoints can be changed at any time during a
•
cycle.
Note: For printing or exporting data please see the Data Management section of this
instruction manual.
Summary of Contents for LYOSTAR II
Page 1: ...LYOPHILIZER OPERATOR S MANUAL FTS SYSTEMS LYOSTAR II STANDARD...
Page 2: ......
Page 6: ...iv Rev 012 04 11 SP Scientific 2011...
Page 29: ...FTS Systems LyoStar II Operating Your Lyophilizer Rev 012 04 11 21 SP Scientific 2011...
Page 38: ...Data Management FTS Systems LyoStar II 30 Rev 012 04 11 SP Scientific 2011...
Page 52: ...Optional Isolation Valve FTS Systems LyoStar II 44 Rev 012 04 11 SP Scientific 2011...
Page 58: ...Optional Cleanroom Configuration FTS Systems LyoStar II 50 Rev 012 04 11 SP Scientific 2011...
Page 66: ...General Maintenance FTS Systems LyoStar II 58 Rev 012 04 11 SP Scientific 2011...
Page 70: ...Appendix A Troubleshooting FTS Systems LyoStar II 62 Rev 012 04 11 SP Scientific 2011...
Page 71: ......