Atlas™ Series Heated Platens
13
prolong the life of operation of both the safety cut out switch (
5
) and
the water cooling blocks (
6
) of the Atlas™ Series Heated Platens
themselves.
If the Atlas™ Series Heated Platens are not going to be used for a long
period of time, the cooling block parts (
6
) of both the top (
1
) and lower
(
2
) heated platen assemblies should be stored
without any water
inside them
. After disconnection of the silicone tubing, any water left
inside can be expelled carefully by connection of either one of the
water connection ports (
3
) to an airline to blow the water out and away
to a drain. (See
Fig 1.
and
Fig 4.)
Using with the Constant Thickness Film Maker Accessory
If the Constant Thickness Film Maker Accessory P/N GS15640 is to be
used with the Atlas™ Series Heated Platens, the cooling chamber
assembly (
7
) provided as part of this Film Maker Accessory should be
included in the water cooling system as shown in
Fig 4
.
Fig 4. Flow Connections for Cooling Block Assembly of Constant
Thickness Film Maker Accessory P/N GS15640
7
6
6
5
2
Water OUT
Water IN