SC5 Signal Conversion
Performance Measurements
•
56
Figure 36: SC5 output noise 0.1 - 10 Hz.
Typical RMS and peak-to-peak noise values in the 0.1 Hz to 10 Hz range are listed below:
RMS noise p-p noise
< 170 nV
< 1.05 µV
Table 11: Typical maximum output noise level in the 0.1 to 10 Hz range.
Noise 10 Hz – 300 kHz
The noise of the SC5 outputs above 10 Hz is measured with the same set-up as described in the previous section. The
upper frequency limit is given by the analog bandwidth of the multimeter.
The following settings have been used for the measurement:
SC5
DMM
Measured output:
AO1
Mode of operation: Digitizer, AC volts
Output amplitude: 0 V
Range:
100 mV
Sample rate:
1.8 MS/s
Resolution:
10 bit
The following graph shows a typical trace for the output noise in the 10 Hz to 300 kHz range: