SC5 Signal Conversion
Performance Measurements
•
51
Analog inputs AC performance measurements
Frequency response
The frequency response of the analog inputs is measured by applying a sinewave with an amplitude of 1 V from a
Nanonis OC4. The amplitude was calibrated using a
National Instruments 4071
PXI digital multimeter operated in
DMM mode.
The following settings have been used for the measurement:
SC5
OC4
Measured input:
AI1
Output amplitude: 1 V
Output range:
1 V
Output waveform: Sinewave
Figure 29: Frequency response of the SC5 analog inputs from 10 Hz to 500 kHz.
Linearity
Linearity is measured by applying a 100 Hz sinewave with amplitudes ranging from 10 V (0 dBFS) to 100 µV (-100
dBFS) generated by AO1 to AI1. The amplitude was calibrated each time using a
National Instruments 4071
PXI
digital multimeter operated in digitizer mode at 5 kS/s. Oversampling of the input is 256×, and the following settings
have been used for the measurement:
SC5
DMM
Measured input:
AI1
Mode of operation: Digitizer, AC volts
Input oversampling: 256×
Range:
10 V and 100 mV
Sampling rate:
5 kS/s
Resolution:
18 bit
Note: The amplitude was determined from the peak to peak value determined from the oscilloscope trace. This is not
a lock-in measurement.
10
10
1
10
2
10
3
10
4
10
5
−3
10
−4
10
−2
10
−1
10
0
10
1
Input voltage [V
]
Frequency [Hz]