Spectrometer
B
=
0
⋅
n
⋅
I
2
R
(15)
Figure 11: Magnetic Field of the Trim Coil
For the PHOIBOS 150 analyzer B[
μ
T] = 0.017 x I[mA] and for the PHOIBOS 100 analyzer
B[
μ
T] = 0.026 x I[mA]. In the case of an earth magnetic field of about 50
μ
T with the
lens axis nearly parallel to the field one needs about 85 mA or -85 mA to compensate
the residual magnetic field within the hemispheres.
To find the optimal value for the current, the signal measured by the analyzer at low
pass energies (< 10 eV) should be maximized by changing the coil current. If the analyz-
er is equipped with a coil (short-circuit of PIN 12 to ground of the analyzer feed-
through) and the power supply HSA 3500 is equipped with a current module, the con-
trol software allows the setting of the coil current. In all programs the coil current set-
ting can be found in the analyzer settings menu as an additional field.
3.4 Slit Orbit Mechanism
The slit orbit mechanism is used to manually change the entrance and exit apertures of
the analyzer. The mechanism positions an entrance and exit opening of the appropriate
size in the analyzer entrance and exit plane. The entrance beam is defined by a pair of
slits which are about 6 mm apart. The posterior slit lies in the entrance plane and
defines the analyzer energy resolution (see equation 5 on page 18), while the anterior
slit serves to match the angular spread for the analyzer. For a given energy resolution
and a given tolerated analysis area and acceptance angle, the largest possible slit
22
Summary of Contents for PHOIBOS 100
Page 1: ...PHOIBOS Hemispherical Energy Analyzer Series PHOIBOS 100 PHOIBOS 150 3 1...
Page 6: ...Table of Contents PHOIBOS...
Page 10: ...Introduction 4 PHOIBOS...
Page 13: ...Electrical Connections Figure 2 Connection Scheme PHOIBOS 7...
Page 14: ...Components and Connections Figure 3 Analyzer Housing PHOIBOS100 8 PHOIBOS...
Page 15: ...Electrical Connections Figure 4 Analyzer Housing PHOIBOS150 PHOIBOS 9...
Page 42: ......
Page 51: ...SpecsLab Hardware and Software Installation PHOIBOS 45...
Page 52: ......
Page 62: ......
Page 78: ...Analyzer Checks Figure 33 Schematics of the 12 pin Analyzer Feedthrough 72 PHOIBOS...
Page 80: ......
Page 86: ......
Page 92: ...List of Figures Figure 39 Alignment Pin 84 II PHOIBOS...
Page 93: ...List of Figures PHOIBOS III...
Page 94: ......
Page 96: ......
Page 98: ...Index PHOIBOS...