Troubleshooting
An arrow (
→
) after a statement indicates a separate troubleshooting procedure either
given here or in another manual.
Possible Cause
perform check, test, or troubleshooting
procedure no.
→
X-Rays off
check the X-ray source and the sample current
No voltage at detector
check detector voltages
Spectrum definition incorrect
check spectrum definition
Cable connection faulty
check cable connections
Preamplifier box defective
→
Counter device or control unit defective contact SPECS
No energy sweep voltage
→
and .
Spectrometer voltages incorrect
check spectrometer voltages section
→
Improper adjustment of slit orbit
check proper adjustment section
→
Table 5: No Spectrum
possible cause
perform check, test, or troubleshooting
procedure no.
→
X-ray intensity too low
check whether on Ag sample photo current
≥
0.2 - 0.3 nA/W
sample is dirty
sputter until C and O peak in the spectrum
disappears
sample too rough
remove roughness
incorrect analyzer - sample distance
adjust distance to 40 mm
lens system and HSA out of focus
check lens and HSA electrodes section
→
CEM yield to low
measure detector supply voltages
→
conversion voltage too low
check conversion voltage
→
section 3.5.4
MCD preamplifier setting changed
→
some channels of the preamplifier unit
defective
check operation of the preamplifier by use of
a separated channel mode of the software or
→
magnetic fields
Check whether the amount of the deviation
depends on the energy of the measured elec-
trons and on the pass energy (better by UPS).
Check the influence of an external permanent
magnet near the vacuum chamber.
physically incorrect adjustments like:
gridsdamaged, lenses misaligned,
Open analyzer system and check. Do this only
if all other kinds of faults can be excluded.
52
Summary of Contents for PHOIBOS 100
Page 1: ...PHOIBOS Hemispherical Energy Analyzer Series PHOIBOS 100 PHOIBOS 150 3 1...
Page 6: ...Table of Contents PHOIBOS...
Page 10: ...Introduction 4 PHOIBOS...
Page 13: ...Electrical Connections Figure 2 Connection Scheme PHOIBOS 7...
Page 14: ...Components and Connections Figure 3 Analyzer Housing PHOIBOS100 8 PHOIBOS...
Page 15: ...Electrical Connections Figure 4 Analyzer Housing PHOIBOS150 PHOIBOS 9...
Page 42: ......
Page 51: ...SpecsLab Hardware and Software Installation PHOIBOS 45...
Page 52: ......
Page 62: ......
Page 78: ...Analyzer Checks Figure 33 Schematics of the 12 pin Analyzer Feedthrough 72 PHOIBOS...
Page 80: ......
Page 86: ......
Page 92: ...List of Figures Figure 39 Alignment Pin 84 II PHOIBOS...
Page 93: ...List of Figures PHOIBOS III...
Page 94: ......
Page 96: ......
Page 98: ...Index PHOIBOS...