Spectrometer
Figure 12: External Rotary Dial for Positioning
The indicators on the rotary dial are used for positioning, but it must be taken into ac-
count that the rotary dial has some backlash. The correct slit positions are defined by
spring loaded indexing balls. Because of the rotary feedthroughs backlash, rotate it
beyond the desired position until the indexing ball snaps in.
After positioning the dial, jog it back and forth to ensure that the index is probably en-
gaged. The correct positioning of the entrance slits can also be checked by looking
through the view port. In a configuration with two exit slits, the positioning of the exit
slit is not critical because of the end stops. In configurations with a third exit opening
please be aware that positioning of slit B can be tricky and require some practice. An
analyzer configured with two exit slits can be reconfigured on request.
Figure 13: Entrance and Exit Slit Rings (Slit Combination 4-B)
24
Summary of Contents for PHOIBOS 100
Page 1: ...PHOIBOS Hemispherical Energy Analyzer Series PHOIBOS 100 PHOIBOS 150 3 1...
Page 6: ...Table of Contents PHOIBOS...
Page 10: ...Introduction 4 PHOIBOS...
Page 13: ...Electrical Connections Figure 2 Connection Scheme PHOIBOS 7...
Page 14: ...Components and Connections Figure 3 Analyzer Housing PHOIBOS100 8 PHOIBOS...
Page 15: ...Electrical Connections Figure 4 Analyzer Housing PHOIBOS150 PHOIBOS 9...
Page 42: ......
Page 51: ...SpecsLab Hardware and Software Installation PHOIBOS 45...
Page 52: ......
Page 62: ......
Page 78: ...Analyzer Checks Figure 33 Schematics of the 12 pin Analyzer Feedthrough 72 PHOIBOS...
Page 80: ......
Page 86: ......
Page 92: ...List of Figures Figure 39 Alignment Pin 84 II PHOIBOS...
Page 93: ...List of Figures PHOIBOS III...
Page 94: ......
Page 96: ......
Page 98: ...Index PHOIBOS...