Calibration
For more detailed information please contact SPECS support.
7.7 Gain Calibration with XPS
See section 8.2.1 , "Check Kinetic Energy Scale" on page 66. Run spectra for the gold
and the copper peak (use clean samples, sputtered). Measure the distance between the
peak maxima of gold and copper. If the value found differs strongly from 848.66 eV
please inform SPECS. You can correct the gain (default value = 1) by
gain
=
distance
848.66
eV
(27)
Note the default voltage range you have used while measuring with the desired meth-
od (see SpecsLab2 menu ’Analyzer/Settings’).