Energy Scale Tests with XPS
Cu L
3
MM
334.94 ± 0.01
567.96 ± 0.02
Table 14: Calibration Binding Energies for non-monochromated Mg K
α
X-rays
8.3 Specification Check
8.3.1 Survey Spectrum of Silver
The XPS performance of an energy analyzer is usually determined using a silver sample.
A clean silver sample is introduced into the vacuum chamber and cleaned by ion sput-
tering. Use the same settings as in the overview spectrum of silver enclosed with the
Specification Report on the analyzer. For example:
Excitation
Mg K
α
X-ray power
300 W
Slit
entrance: largest slit / exit: open
Mode
Medium Area
Binding energy range 1000 - 0 eV
Energy step
~300 meV
Dwell Time
100 ms
Pass energy
~13 eV
A typical XPS overview spectrum taken with a PHOIBOS analyzer on silver is shown be-
low (figure 31).
67
Summary of Contents for PHOIBOS 100
Page 1: ...PHOIBOS Hemispherical Energy Analyzer Series PHOIBOS 100 PHOIBOS 150 3 1...
Page 6: ...Table of Contents PHOIBOS...
Page 10: ...Introduction 4 PHOIBOS...
Page 13: ...Electrical Connections Figure 2 Connection Scheme PHOIBOS 7...
Page 14: ...Components and Connections Figure 3 Analyzer Housing PHOIBOS100 8 PHOIBOS...
Page 15: ...Electrical Connections Figure 4 Analyzer Housing PHOIBOS150 PHOIBOS 9...
Page 42: ......
Page 51: ...SpecsLab Hardware and Software Installation PHOIBOS 45...
Page 52: ......
Page 62: ......
Page 78: ...Analyzer Checks Figure 33 Schematics of the 12 pin Analyzer Feedthrough 72 PHOIBOS...
Page 80: ......
Page 86: ......
Page 92: ...List of Figures Figure 39 Alignment Pin 84 II PHOIBOS...
Page 93: ...List of Figures PHOIBOS III...
Page 94: ......
Page 96: ......
Page 98: ...Index PHOIBOS...