Analyzer Checks
Figure 31: XPS on Silver, Wide Scan (PHOIBOS 150 MCD9)
Parameters for the survey spectrum in figure 33:
Lens mode:
Medium Area
Slit:
entrance: largest slit / exit: open
Sample current:
160 nA at 300W Mg K
α
8.3.2 Intensity and Resolution
Use the same settings as in the Ag spectra enclosed with the Specification Report on the
analyzer.
For example:
Excitation
Mg K
α
X-ray power
300 W
Slit
entrance: largest slit / exit: open
Mode
Medium Area
Bin. energy range
378 - 364 eV
Dwell Time
100 msec
No. of scans
1
Step width
25 - 30 meV
Pass energy
8-9 eV
A well resolved Ag 3d doublet is shown in figure 32. This spectrum is typical for a
PHOIBOS.
68