Table 2: Main X-ray source excitation lines
Table 3 summarizes the energy shifts and relative intensities of satellite lines of the main excit-
ations.
Line
Energy shift
Mg (eV)
Relative intensity
Mg (%)
Energy shift Al
(eV)
Relative intensity
Al (%)
K
α3
8.4
8.0
9.8
6.4
K
α4
10.1
4.1
11.8
3.2
K
α5
17.6
0.6
20.1
0.4
K
α6
20.6
0.5
23.4
0.3
K
β
48.7
0.5
69.7
0.6
Table 3: Mg/ Al K
α
satellites
6.1.3
Cross-Talk
There is a small rod behind the Al-window which reduces the cross-talk between the excitations
of the two anodes to below 0.35 %. The rod is oriented parallel to the groove at the anode top
which separates the two anode sides from each other.
Figure 21: Cross-talk suppression
A cross-talk spectrum is shown in Figure 22. The peak highlighted in the aluminium spectrum is
due to Mg cross-talk.