Figure 10: Activating a preset
6. Press the
Activate
softkey. This starts the following actions:
Current parameter values are replaced by those from the preset.
The water cooling will start, if it is not already running. A message will appear on the
screen as it starts.
The display switches to the operating screen.
The filament current ramps up to the standby current.
The high voltage is ramped up to the value in the preset.
The filament current is ramped up until the emission current enables the target power
(power = high voltage × emission current).
3.2
Adjusting Settings During Operation
After starting the source, the operating menu is displayed. The top parameter shows the
power. This is the high voltage multiplied by the emission current. You can adjust the power
with the encoder wheel.
Figure 11: Operating menu