Figure 18: Anode water cooling flange and tube
After fitting the cooling tube, blow into one of the ports to make sure that air flows through. The
pressure required should be similar to that needed to blow up a balloon.
5.5
Replacing the Aluminium Window
The aluminium radiation window (standard thickness 2 µm) suppresses Bremsstrahlung and
prevents the input of stray electrons generated by the filament from entering the energy ana-
lyzer. Additionally, the X-ray window acts as a barrier between the main chamber and anode
volumes.
The Al window can be torn, perforated or otherwise destroyed by a significant pressure dif-
ference between the source interior and chamber.
When replacing the Al window, it is necessary to remove the entire X-ray source from the
vacuum system.
Danger—High Voltages Present!
It will take 3 minutes before all high voltages are absent in the system. Wait
at least 3 minutes before disconnecting any cables from the power supply or
the X-ray source.
To replace the aluminium window: