Spectracom Corporation
EPSILON CLOCK MODEL EC22S
EPSILON CLOCK User’s Manual
4-7
4.2.2.6
Reset IP address screen
Reset IP address:
IP: 192.168.0.100
[Reset]
[Back]
This function may be useful in case the equipment is in DHCP mode and connected to a
network with no DHCP server. Executing "Reset" programs the EC22S to the default
192.168.0.100 IP address. You can then connect a PC with a crossed RJ45 cable and modify
the IP address using the web page (System setup>Network setup).
4.2.2.7
Extraction screen
Slave on Right Side
Safe to be extracted
-Halt-
[Back]
The Extraction screen relates to the extraction of a Clock module for repair and guides the
operator through the process.
For an extraction, be careful to:
•
Extract the Slave Clock only. If a Master Clock is extracted while distributing, a glitch may
occur on the distributed signals
•
Switch off the AC power on the side of the Clock module to be extracted. While it will not
harm the equipment if done incorrectly, repeated removal with AC power applied may
damage the hardware over time.
The first message displays the status: the Slave Clock is safe to be extracted. If the Clock that
will be extracted is designated as the Master, go to the "Distribution mode" setup screen and
select "Forced" for the Clock that will remain in the equipment.
Go back to this screen and proceed with [Halt] to stop the module software. Move the cursor to
the “Halt” field. Click the "enter" key. A "Halt in progress" message is shown during software
shut-down operations The “Halt” screen is then displayed.