Getting Started
Spectral Products
o
time delay
sets the time in ms to wait before
echo back.
o
port number
sets the com port of the
monochromator.
o
ACTUAL
λ/
PEAK
λ:
Enter the measured
wavelength of the peak from the known source.
o
DESIRED
λ:
Enter the theoretical peak
wavelength of the known source.
o
time out, echo byte, status byte, and cancel
byte
are diagnostic parameters associated with
the LabVIEW program.
o
Run
enters the values and recalibrates with the
current wavelength
.
•
Calibrate
allows recalibration of the monochromator positioning scale factor. A source of known
wavelength is needed for recalibration. The user must first execute a scan and note the displayed
peak value.
Calibrate
should always be used immediately after using the ZERO command.
CM-Series
DK Series
o
Exit
closes the Command window.
•
Echo
verifies communications with the monochromator.
•
Flush
cleans the serial communications receiver buffer.
•
Goto
moves the monochromator to a selected position.
•
Reset
returns the grating to home position.
Spectral
Products
2659 Pan American Fwy, Suite A, NE Albuquerque, NM 87107 Ph. (505) 343-9700 Fax (505) 343-9705
12
•
Select
selects the grating that will be used.