DI CARTRIDGE REPLACEMENT:
1. When the reading on the TDS meter (set to “Line 3”) displays “001”, it is
time to replace the SilicaBuster™ Extended Life Super DI cartridge.
Part number DI-SB-10HC.
2. Turn the system off and remove the cartridge housing by rotating it to
the left until it is free of the threaded head on the bracket.
3. Make sure the DI cartridge is installed in the correct direction as marked on
the cartridge shell and be sure that the top seal is securely attached to the
top of the cartridge.
4. Tighten the cartridge housing by rotating it to the right and hand tighten.
5. Turn on system and check for leaks.
TRM-1 METER Specifications:
Range
0 -9990 PPM
Resolution
1 PPM (0 -999 PPM)
10 PPM (1000 -9990 PPM)
Accuracy
± 2%
Conversion Factor
NaCL
Power Source
(2) 1.5V button batteries (LR44 or equiv.)
Battery life
Approx 1000 hours