MAXCAP DI MAINTENANCE AND REPLACEMENT:
Procedure:
1. When the reading on the right-hand TDS meter (set to “IN”) displays 50% of
the reading on the left-hand TDS meter (set to “OUT”), it is time to replace
the
MAXCAP
®
DI cartridge.
2. Make sure the DI cartridge is installed in the correct direction as marked on
the cartridge shell and be sure that the top seal is securely attached to the
top of the cartridge.
3. Tighten the cartridge housing by rotating it clockwise and hand tighten.
4. Turn on system and check for leaks.
SILICABUSTER MAINTENANCE AND REPLACEMENT:
Procedure:
1. When the reading on the right-hand TDS meter (set to “OUT”) displays
“001”, it is time to replace the SilicaBuster™ DI cartridge.
2. Make sure the DI cartridge is installed in the correct direction as marked on
the cartridge shell and be sure that the top seal is securely attached to the
top of the cartridge.
3. Tighten the cartridge housing by rotating it clockwise and hand tighten.
4. Turn on system and check for leaks.
DM-1 METER Specifications:
Range
0-1999 PPM
Resolution
1PPM (1-999 PPM)
Accuracy
2%
Probe
0.25 inches
Power Source
(2) 1.5V button batteries (A357 or equiv.)
Battery life
Approx 1000 hours