Metal Clip
Mounting
Bracket
Adjusting Pin
Strap
Bracket
ATTACHING THE LAMP TO THE HEAD STRAP
Adjust the head strap to
fi
t your head or helmet. The adjusting pin on the strap’s mounting
bracket has been factory-tightened to hold the lamp
fi
rmly in position. Use an Allen wrench
to adjust the pin further, if necessary. To attach the lamp to the strap, pull the mounting
bracket away from the strap bracket. Pull back the metal clip on the mounting bracket and
slide the lamp down onto the bracket. Release the metal clip to secure the lamp. Place the
strap and lamp on the helmet, securing the strap under its back lip. Position the lamp by
moving it forward and back until the lights are focused on the work surface.
UV INTENSITY VERIFICATION
To ensure that your EagleEye EE-365 lamp is operating at the required UV intensity, it
should be checked periodically. The AccuMAX
TM
XRP-3000 radiometer/photometer kit uses
a dual-wavelength UV-A/VIS sensor detector to accurately measure ultraviolet irradiance
or visible light. The AccuMAX XR-1000 digital readout unit with the XS-365 UV sensor or
the Spectroline DM-365XA digital readout unit are also recommended for accurate UV-A
measurement. The meters are specially designed to measure the UV irradiance from
320–400nm with a peak at 365nm.
For more information about these units, please contact the Customer Service Department
at Spectronics Corporation. In the U.S. and Canada, call toll-free 1 800-274-8888. Outside
the U.S., call 516-333-4840.
CARE AND USE OF LITHIUM-ION (Li-ion) BATTERIES
• Do not allow Li-ion batteries to fully discharge frequently, since this will put a strain on the
batteries. They will work more ef
fi
ciently on multiple partial discharges through regular use
(every 2-3 weeks), and frequent recharges.
• The batteries may be left in the battery cradle after charging. The charger automatically cuts
the charge when the charge is full.
• Remove the battery from the lamp and store it separately.
Before prolonged storage (30 days
or more), charge the battery for at least two hours
. Store partially charged for best results.
• Store in a fairly cool environment away from sunlight, heat and humidity. Store the battery at
temperatures between 5 °C and 20 °C (41 °F and 68 °F).
NOTE:
The battery self-discharges
during storage. Higher temperatures (above 20 °C or 68 °F) will reduce the battery storage life.
• Lithium-Ion batteries are subject to disposal and recycling regulations that vary by country and
region. Always check and follow your applicable regulations before disposing of any battery.
ENVIRONMENTAL SPECIFICATIONS
The EagleEye EE-365 lamp is designed to be safe under the following conditions:
• Indoor use;
• Altitude up to 2,000 m (6,562 ft.);
• Temperature 5°C to 40°C (41°F to 104°F);
• Maximum relative humidity 80% for temperatures up to 31°C (88°F) decreasing linearly
to 50% relative humidity at 40°C (104°F);
• Mains supply voltage
fl
uctuations not to exceed ±10% of the nominal voltage;
• Installation Category II;
• Pollution Degree 2.