118159-001 REV. A
2 OF 7
Preset Mode
The eSL features a Preset Mode that allows for easy setting of the kV and mA program values. While the
power supply is in Standby/HV OFF depress and hold in the green front panel HV OFF button. While
doing so, the actual kV and mA values programmed into the power supply will be displayed on the kV
and mA meters. Rotate the kV encoder to set the kV program set point. Rotate the mA encoder to set
the mA program set point. The two encoders are setup so if you turn the dial slowly they will change by
the smallest increment possible. If you rotate the encoder quickly they will change with larger
increments. Release the HV OFF button and the eSL will go back to the Standby/HV OFF mode.
Main Menu Screen
Pushing the current encoder once will bring you to the main menu screen which displays the following:
Local/Remote Control
Features
Tutorial
Diagnostic
Exit
Local/Remote Control
Features