118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
57
4.4.30
Request Current X-Ray Tube Anode Heat Loading
Description:
The host requests that the firmware report the current anode heat
load of the X-Ray tube as a percentage of the allowed maximum for
the tube.
Direction:
Host to supply
Syntax:
<STX><77><,><CSUM><ETX>
Response:
<STX><77><,><ARG><,><CSUM><ETX>
Where:
<ARG> = 0 - 100 in ASCII format, representing 0% (no heat), to
100% (maximum allowed heat).
Example:
<STX>77,<CSUM><ETX>