118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
65
4.5.4
Set Exposure Settings Max Exposure Time
Description:
The host requests that the firmware change the maximum exposure
time. Note: The maximum exposure time is the maximum time you
can set using the “Set exposure” settings” command (command =
50), and the “Set exposure Time” command (command = 72). It
does not change the current exposure time set by command 50 or
72. This is a service command and should be used with caution
because it over-rides the tube table and allows exposure times that
may damage the supply or X-Ray tube. This command should not
be used in general operation.
Direction:
Host to supply
Syntax:
<STX><40><,><ARG><,><CSUM><ETX>
Where:
<ARG> = 0 - 20000 in ASCII format, where <ARG> is the time in
milliseconds.
Example:
<STX>40,5000,<CSUM><ETX>
Response:
<STX><40><,><$><,><CSUM><ETX>