118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
68
4.5.7 Set Ready For Exposure Time
Description:
The host requests that the firmware change the ready for exposure
time. This is the amount of time after entering Prep before X-Ray
On is allowed. Values of less than 2 seconds will result in improper
operation. This command is intended for system testing only.
Direction:
Host to supply
Syntax:
<STX><34><,><ARG><,><CSUM><ETX>
Where:
<ARG> = 0 - 5000 in ASCII format (time in milliseconds)
Example:
<STX>34,4095,<CSUM><ETX>
Response:
<STX><34><,><$><,><CSUM><ETX>
Note: The PMX unit has a maximum prep time of 30 seconds,
meaning that no matter what you program. If the total time in prep
(Prep + Exposure control on time) exceeds 30 seconds the
exposure will be terminated. Increasing the Ready for Exposure
time reduces the amount of time you have to actually make the
exposure.