118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
86
In Smart AEC
Mode
Start 2-Point
set exposure
Get Programmed kV, filament
and mAs
Set:
Proposed Speed = Low Speed
Get programmed kV, filament
and mAs
Set:
Proposed Speed = High Speed
No
Yes
Go to Tube Table and find
highest allowed mA setting for
current combination of
kV/Filament/Speed
Settings.
This is proposed mA = mA1
Divide set exposure mAs by
proposed mA (mA1) to get
proposed exposure time “T1”
Go to Tube Table and find
maximum exposure time for
current kV/mA/Filament/speed
Setting (T-Max)
T1 <= T-Max
(Settings Valid)
Set Result code to Ok
Set Exposure
kV = programmed kV
mA = mA1
Exposure Time = T1
Filament = Proposed Fil
Speed = Proposed Speed
Proposed Speed =
Low?
Change Proposed
Speed to High
Proposed
mA1 > 20 mA
Change Proposed
mA1 to 10 mA lower
(mA1 = mA1 – 10 mA)
Set Result Code to Bad mAs
Value. Do not update
exposure mA/Filament/
Speed settings
Try new
kV/Filament/Speed
Combination
Y
No – Settings not valid
N
Y
Try new
mA1 setting
Yes
Settings Valid
Done
N
2 Point Mode
kV & mAs to exposure
settings
Fig 17