Table 2 – Guidance and MANUFACTURER’S declaration – ELECTROMAGNETIC EMISSIONS – for all ME
EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The PMX is intended for use in the electromagnetic environment specified below. The customer or the
user of the PMX should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment
–
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV for power
supply lines
±
1 kV for
input/output
lines
±
2 kV for power
supply lines
±
1 kV for
input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±
1 kV line(s) to
line(s)
±
2 kV line(s) to
earth
±
1 kV line(s) to
line(s)
±
2 kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions
and
voltage
variations
on power supply
input lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T)
for 5 s
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T)
for 5 s
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the PMX
requires continued operation during
power mains interruptions, it is
recommended that the PMX be powered
from an uninterruptible power supply or a
battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE
U
T is the A.C. mains voltage prior to application of the test level.