118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
41
4.4.15
Request kV Set-Point
Description:
The host requests that the firmware report the kV set point. This is
the current value of the exposure kV setting. The value is 0-4095
scaled the same as the kV program value: 50kV/4095 = 12.21V/Bit
Direction:
Host to supply
Syntax:
<STX><14><,><CSUM><ETX>
Response:
<STX><14><,><ARG><,><CSUM><ETX>
Where:
<ARG> = 0 - 4095 in ASCII format, scaling = 12.21V/Bit
Example:
<STX>14,<CSUM><ETX>