118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
50
4.4.24
Get Break After Exposure Hang Time
Description:
The host requests that the firmware report the current Break After
Exposure Hang Time. This is the amount of time the X-Ray tube rotor is
driven after the end of an exposure. The time is specified in milliseconds.
See the Set Break After Exposure Command for more details. Also see
Section 7.0 for more details on Break After Exposure operation.
Direction:
Host to supply
Syntax:
<STX><37><,><CSUM><ETX>
Response:
<STX><37><,><ARG><,><CSUM><ETX>
Where:
<ARG> = 0 - 65535 in ASCII format. This is the Break After Exposure
Hang time in milliseconds.
Example:
<STX>37,<CSUM><ETX>