SL2KW SERIES MANUAL
12
118088-001 REV C
It is recommended that analog signals be isolated via
isolation amplifiers.
All cables should be shielded
with the shield being returned to the chassis ground
of the High Voltage Power Supply.
Voltage Source 0-10V = 0-100% of Rated Output
Z in = 10M ohm
Remove jumpers for
remote programing
10
11
9
Z in = 10M ohm
J5
3
7
8
6
5
4
2
1
-
+
V
Remote
Current
Program
Remote
Voltage
Program
+
V
-
PS Common
Figure 3.2A – Wiring Diagram for Remote Programming via Voltage
Fully Counterclockwise = Zero Output
Fully Clockwise = 100% of Rated Output
All cables should be shielded with the shields
being returned to the chassis ground of the
High Voltage Power Supply.
If connections to external potentiometers are
excessively long (>10ft), motorized
potentiometers are recommended.
Remote
Current
Control
Remove jumpers for
remote programing
11
10
8
9
7
5
6
4
Z in = 10M ohm
+10.32V
Z in = 10M ohm
1mA max.
20k ohm
CW
CW
J5
1
3
2
20k ohm
PS Common
Remote
Voltage
Control
Figure 3.2B – Wiring Diagram for Remote Programming via External