background image

81233-4r1 

 

Page 3 of 9 

Change History 
 

Section 

Reason for Change 

Issue 

All 

Original 





Added regulatory specifications 
Added environmental conditions 
Added mechanical 
Added input and output connections details 
Added setup and adjustments 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 

 

Contents 

1

 

Unit Description ........................................................................................................................................ 4

 

2

 

Safety ......................................................................................................................................................... 4

 

3

 

Regulatory Specifications ....................................................................................................................... 4

 

4

 

Environmental conditions ....................................................................................................................... 5

 

5

 

Mechanical ................................................................................................................................................ 5

 

6

 

Installation ................................................................................................................................................. 6

 

7

 

Input and Output Connections ................................................................................................................ 7

 

8

 

Set Up & Adjustments .............................................................................................................................. 9

 

 

 
 

 

Summary of Contents for XMPG10P10

Page 1: ...ION INSTRUCTIONS Document Number 81233 4 Issue A 1 Date 23 01 18 16 04 2021 Issuing Authority I 5650 10465C Checked Approved Tel 44 0 1798 873986 Fax 44 0 1798 872479 E mail hvsales spellmanhv co uk Broomers Park Broomers Hill Lane Pulborough W Sussex RH20 2RY ...

Page 2: ...quipment is turned off and the capacitance of both the load and power supply are grounded Allow adequate time for discharge of internal capacitance of the power supply Do not ground yourself or work under wet or damp conditions Servicing Safety Maintenance may require removing the Instrument cover with the power on Servicing should only be done by qualified personnel aware of the hazards If in dou...

Page 3: ...fications Added environmental conditions Added mechanical Added input and output connections details Added setup and adjustments 1 Contents 1 Unit Description 4 2 Safety 4 3 Regulatory Specifications 4 4 Environmental conditions 5 5 Mechanical 5 6 Installation 6 7 Input and Output Connections 7 8 Set Up Adjustments 9 ...

Page 4: ...for use in a Pollution Degree 2 Installation Category II environment Consideration should be given to conducting the following tests with the unit installed in the end product Dielectric Voltage Withstand Test between live parts of the unit and the end product chassis Permissible Limits Tests with the unit installed in the end product Temperatures on power electronic components transformer winding...

Page 5: ...te humidity maximum 25 gm Maximum storage period 0 25 year at above conditions Long term storage 3 year special packaging required 5 Mechanical 5 1 Mechanical outline 185mm x 105mm x 33 5mm The mass of the module is nominally 2kg 105 0 33 5 185 0 66 5 9 7 30 0 23 7 46 0 14 7 15w D plug HOLE Ø4 5mm IN 3 POSITIONS MOUNTING HOLE TEMPLATE VIEWED FROM THE TOP OF UNIT NOTE Max thread depth into unit is ...

Page 6: ...gree 2 Installation Category II environment The input and output connectors are not intended for field connections and should only be connected to internal wiring in the end product The unit is intended for use as a component and no surface of the unit should be accessible in the end product 6 3 Electrical Installation The unit must be terminated safely before operation Hazardous voltages will be ...

Page 7: ...r 7 1 4 HV voltage program input 0 10V input corresponding to 0 10kV output Accuracy 2 of full scale Zin 5MΩ 7 1 5 HV voltage monitor 0 10V output corresponding to 0 10kV output Accuracy 2 of full scale Zout 10kΩ 7 1 6 Emission current monitor 0 10V output corresponding to 0 1mA emission current Accuracy 3 of full scale Zout 10kΩ 7 1 7 Filament current monitor 0 10V output corresponding to 0 5A fi...

Page 8: ... 0 10V output 11 Filament current monitor 0 10V output 12 Emission current program input 0 10V input 13 Emission current program output 0 10V from internal pre set 14 Signal ground 15 Pre set maximum filament current set value 0 10V from internal pre set The filament Preheat level and current limit are set by internal presets accessible through the case side If external high voltage control is not...

Page 9: ... heat current 0 5V corresponds to 0 5A 2 5A filament current 8 6 If external HV control is required connect the control signal to pin 8 with return to pin 14 0 10V corresponds to 0 10kV 8 7 If internal HV control is required link pins 7 and 8 then turn the HV preset adjustment on the case front whilst monitoring the voltage at pin 8 return to pin 14 Set for the HV potential required 8 8 If externa...

Reviews: