Chapter 11- Calibration method
Frequency: calibrate once a month.
Calibration methods: via clinical thermometers (such as the axillary mer-
cury thermometer) measurement the temperature, to compare the result
testing by this machine. If the deviation is too large, it can be offset com-
pensation key “F4” to fix the correct value.
Note:
If the measurement results is large different from the temperature tested
by the clinical thermometers, the machine should immediately be cali-
brated.
Chapter 12- Cautions
1. The protective lens of probe is the most damageable part of the infrared
thermometer. Hence, it must be taken good care.
2. Cleaning method of probe lens: use the cotton bud or soft cloth with
water or alcohol to wipe the lens lightly.
3. Never charge the batteries or throw any of them into fire. Please discard
Temp’O Laser
ed
rt-
rs,
to
or
Symptom
of machine failure
Key trigger invalid, no
response
No display or display lack
of strokes
No voice prompts
Press measuring key is
invalid, no response
Solution
measurement.
Note that the effective
measurement of dis-
tance is: 5 to 10cm
Do not self-demolish
thermometer
Do not self-demolish
thermometer
See whether the voice
prompts
switch
is
turned on, when the
sound identification: “1”
is turn on the sound, “0”
is off the sound.
1. Low battery, please
replace the new bat-
tery;
2. Whether the battery
polarity is correctly in-
stalled.
Other method
Same as above
Same as above
Same as above
Same as above