MAINTENANCE
pH Probe
It is important to keep the pH probe wet when not in use. The
probe is protected by a plastic bottle containing solution. To
use or store the probe:
1. Rotate the bottle to remove the bottle from the probe. Pull
down the cover and remove it from the probe.
2. After use, put the cover back on the probe and plug the probe
into the bottle. Rotate the bottle to fit into the cover tightly.
The following actions will keep the probe in good
working condition:
• Always keep the pH glass bulb wet by using the plastic bottle
to protect and store the probe. You can also store it in a KCl
solution. Never use distilled or de-ionized water for storage
• Always rinse the pH probe in de-ionized water before using
• Never touch or rub the glass bulb tip.
ORP Probe
Before using, remove the soaking bottle, soak the probe in
distilled water, and rinse. Gently dry the sensing element.
MAINTENANCE
pH Probe
It is important to keep the pH probe
wet when not in use. The probe is
protected by a plastic bottle
containing solution. To use or store
the probe:
1.
Rotate the bottle to remove the
bottle from the probe. Pull down
the cover and remove it from the
probe.
2. After use, put the cover back on
the probe and plug the probe into
the bottle. Rotate the bottle to fit
into the cover tightly.