19
SETUP MODE
Note…
When using Manual Temperature Compensation Mode (MTC), you must
set the temperature solution in P5.3.
7. At P5.1, press
twice to enter P5.3. The default “25.0” flashes on
the middle of the LCD and P5.3 will appear on the lower portion. To
adjust the solution temperature setting, press ▲ or ▼.
8. Press
to confirm and enter P5.4.
Note…
When using the TDS measurement mode, you must set the TDS
conversion factor in P5.4.
9. At P5.1, press
three times to enter P5.4. The default “0.500”
flashes on the middle of the LCD and P5.4 will appear on the lower
portion. If the TDS conversion factor of the solution is not 0.5, press ▲
or ▼ to adjust the value. Press
to confirm and return to P5.0.