BATTERY AND FUSE REPLACEMENT
1. If “ 

 

” appears in the display, it is indicating that the battery should be replaced. 

2. Turn the meter on it’s face and remove the 2 screws. 
3. Remove the back panel and locate the battery or fuse.
4. Replace the battery or fuse, re-attach the back cover and re-install the screws.

CAUTION
Before attempting to open the case, be sure to disconnect the test leads to avoid electrical shock.

ACCESSORIES
 • Operator’s instruction manual
 • Set of test leads
 • 9-volt battery, NEDA 1604 6F22 type.
 • Fuse (500mA/250V)

TECHNICAL SPECIFICATIONS

DC VOLTAGE

RANGE

RESOLUTION ACCURACY

200mV

100μV

±(0.5% of rdg + 3D)

2000mV

1mV

±(0.8% of rdg + 2D)

20V

10mV

200V

100mV

1000V

1V

±(1.0% of rdg + 2D)

OVERLOAD PROTECTION: 220V rms AC for 200mV range 
and 1000V DC or 750V rms for all ranges.

AC VOLTAGE

RANGE

RESOLUTION ACCURACY

200V

100mV

±(1.2% of rdg +10D)

750V

1V

RESPONSE: Average responding, calibrated in rms of a sine wave.
FREQUENCY RANGE: 45Hz ~ 450Hz 
OVERLOAD PROTECTION: 1000V DC or 750V rms for all ranges.

DC CURRENT

RANGE

RESOLUTION ACCURACY

200μA

0.1μA

±(1.8% of rdg +2D)

2000μA

1μA

20mA

10μA

200mA

100μA

±(2.0% of rdg +2D)

10A

10mA

±(2.0% of rdg +10D)

OVERLOAD PROTECTION: 500mA 250V fuse (10A range unfused).
MEASURING VOLTAGE DROP: 200mV

RESISTANCE

RANGE

RESOLUTION ACCURACY

200

Ω

100M

Ω

±(1.0% of rdg +10D)

2000

Ω

1

Ω

±(1.0% of rdg +4D)

20K

Ω

10

Ω

200K

Ω

100

Ω

2000K

Ω

1K

Ω

MAXIMUM OPEN CIRCUIT VOLTAGE: 3.2V.
OVERLOAD PROTECTION: 15 seconds maximum 220Vrms.