12 WWW.SPEXSEATING.COM

SPEX HEAD SUPPORT USER MANUAL

 WARNING

1.  The distance between the top of the user’s shoulder and the top of 

the back should not exceed 6.5” (16.5 cm).

2.  The Spex Seating System must only be used with wheelchair bases 

that are tested as part of a wheelchair system that conforms to 

performance requirements of ISO 7176-19 and RESNA WC-4:2012, 

Section 19. 

3.  The Spex Seating System should only be used on wheelchair frames 

that provide four securement points and pelvic-belt anchor points 

that conform to the requirements of RESNA WC-4: 2012, Section 19. 

4.  The wheelchair must be labelled as appropriate for use as a 

seat in a motor vehicle, dynamically tested to the performance 

requirements of ISO 7176-19 and installed, used and maintained 

according to the manufacturer’s instructions.

5.  The Spex Seating System must only be used as indicated in this 

user manual.

6.  Both pelvic and shoulder-belt restraints as a part of a complete 

Wheelchair Tie down and Occupant Restraint System (WTORS) 

which meets ISO 10542-1 shall be used to achieve effective 

occupant restraint and optimum protection in a vehicle impact; the 

belt restraints should be compatible with the specific wheelchair, 

and fitted to the user in accordance with WTORS manufacturer’s 

instructions.

7.  Back supports with adjustable recline angles should not be 

tilted back more than 30° beyond the vertical during travel in 

vehicles unless necessary for the medical and postural needs of 

the occupant.

8.  If the back support must be adjusted to an angle that is greater 

than 30° beyond the vertical during travel, the upper shoulder-belt 

anchor point should be adjusted rearward to maintain belt contact 

with the wheelchair occupant’s shoulder and chest. 

Summary of Contents for Head Support

Page 1: ...assed on to the user of this product Please do not remove this user manual before delivery to the end user SUPPLIER This manual must be given to the user of this product USER CAREGIVER Before using th...

Page 2: ...rt is appropriate for the user The intended user of the Spex Head Support is a user with postural support requirements Contents Introduction 2 Safety Instructions 3 Weight Capacity 3 Environment 3 Com...

Page 3: ...ss of our products Our claims relate to the ability of the products to provide safe and reliable positioning and seating on the equipment onto which they are installed Weight Capacity Head Support max...

Page 4: ...The Head Support must not be modified or changed as this voids warranty 4 Only use genuine Spex accessories components and hardware 5 Never lift the wheelchair by the Head Support 6 Never use the Head...

Page 5: ...for signs of stress or cracking Upholstery covers Covers must be replaced regardless of condition Hardware and components Ensure the mechanism components are in place and in safe working order If any...

Page 6: ...cket 2 Partly tighten the bolts on the top clamp so the headrest pad can move but is not loose 3 Position the headrest pad correctly then securely tighten the bolts using a 5mm hex key Adjusting Stem...

Page 7: ...te 2 Loosen the bolts in the positioning mechanism using a 5mm hex key 3 Move the Head Support Pad to the correct position and securely tighten all bolts WARNING Check any links that stick out from th...

Page 8: ...Close the cover replace the lateral wing and tighten the nuts 2 Adjust the width between the lateral pads by loosening the 2 nuts on the back of the centre pad with a 10mm spanner Move the laterals t...

Page 9: ...Wipe clean with mild detergent and water Use disinfectant wipes Medifab recommends using AHP Accelerated Hydrogen Peroxide disinfectant wipes The disinfection product must be used as per the manufact...

Page 10: ...ified and not recommended for transport in a vehicle Eg Vigour Lo and Vigour Mid back supports Assembly Use Maintenance Limitations Adjustments to the set up of the wheelchair must only be carried out...

Page 11: ...located near the hip opposite the shoulder over which the diagonal belt crosses and not near the midline of the occupant Belt restraints should not be routed outside of the wheelchair wheels or over t...

Page 12: ...rding to the manufacturer s instructions 5 The Spex Seating System must only be used as indicated in this user manual 6 Both pelvic and shoulder belt restraints as a part of a complete Wheelchair Tie...

Page 13: ...straint use and Have energy absorbing padding placed between the back edge of the tray and the wheelchair occupant 11 Alterations or substitutions must not be made to the seating system structure part...

Page 14: ...ansported while occupied the wheelchair must still be secured to the WTORS 18 Children whose mass is less than 48 5 lbs 22 kg should be transferred from their seating system into appropriate child res...

Page 15: ...the original date of purchase Medifab will at its option repair or replace it without charge This warranty does not apply to general wear and use including punctures tears or burns nor to the removabl...

Page 16: ...ax 64 3 307 2820 fasterresults spexseating com www spexseating com 32 Detroit Drive Rolleston 7675 New Zealand US REP Medicept 200 Homer Ave Ashland MA 01721 United States Tel 1 508 231 8842 EC REP BE...

Reviews: