DLS 400E3 Operating Manual

Spirent Communications - Page

 

4-7

7104000537  03/04 -2

3)  A check to see if the controller card is functional

Following the self-tests, the simulator re-establishes the loop that was in use before the unit was turned off or
reset. Any impairment modules in the system is set so that all impairments are OFF.

4.7

Common Command Set

As specified in the IEEE 488.2 standard, a number of common commands are required to set up and control
of the standard functions of remote-controlled devices. They can be used with both the IEEE 488 and the
RS–232 interfaces. These common commands are as follows:

*CLS

Clear Status Command

Type:

Status command

Function:

Clears the Event Status Register (ESR). Clearing the Event Status Register also clears 
ESB, the bit 5 of the Status Byte Register (STB). It has no effect on the output queue (bit 4 
of the STB).

*ESE <NRf>

Event Status Enable

Type:

Status command

Function:

Sets the Event Status Enable Register (ESER) using an integer value from 0 to 255, repre-
senting a sum of the bits in the following bit map:

Bits 7 to 0 have values of 128, 64, 32, 16, 8, 4, 2 and 1, respectively. For example, if bits 3 
and 5 are set then the integer value is 40 (8+32).

The ESER masks which bits are enabled in the Event Status Register (ESR).

On power-on, the register is cleared if the Power-on Status Clear bit is 1, or restored if the 
bit is 0 (see *PSC for more details).

*ESE?

Event Status Enable Query

Type:

Status command

Function:

An integer value between 0 and 255 representing the value of the Event Status Enable 
Register (ESER) is placed in the output queue. The possible values are described in the 
*ESE command section.

*ESR?

Event Status Register Query

Type:

Status command

Function:

An integer value between 0 and 255 representing the value of the Event Status Register 
(ESR) is placed in the output queue. Once the value is placed in the output queue, the reg-
ister is cleared. The command turns the REMOTE LED green if the LED was red. The 
possible values are described in the *ESE command section.

Summary of Contents for DLS 400E3

Page 1: ...Operating Manual DLS 400E3 ADSL European Wireline Simulator Revision 2 March 2004...

Page 2: ......

Page 3: ...e IEEE 488 Port GPIB 2 5 2 5 7 2 Connecting the Computer via the Serial Port RS 232 2 6 3 DLS 400E3 CONTROL SOFTWARE 3 1 3 1 About the Software 3 1 3 1 1 Computer Hardware and Software Requirements 3...

Page 4: ...Example using the RS 232 Interface 6 6 6 3 Wireline Simulator Bypass 6 6 6 4 Setting the Relay Refresh and Micro Interrupt Features 6 7 6 4 1 Synchronization of Commands 6 7 6 4 2 Sub routines for Com...

Page 5: ...a simple upgrade to an existing DLS 400E2 the DLS 400E3 offers the highest accu racy for testing European ADSL ADSL2 devices The complete DLS E3PKG solution from Spirent Communications combines the DL...

Page 6: ...se If you already use a DLS 400E2 Wireline Simulator there are upgrades available for DLS 400E3 Please contact your local Spirent Communications sales representative for pricing See Chapter 8 CUSTOMER...

Page 7: ...tool is recommended for use on a regular basis prior to any extensive tests that must be performed Real world noise has a dramatic impact on the performance of ADSL modems and DSLAMs ETSI TS 101 388 h...

Page 8: ...simulator The remainder of this manual provides information about the various aspects of the wireline sim ulator such as loop configurations remote control warranty specifications and contact informa...

Page 9: ...5200E3 Noise Generation System Operating Manual DLS 1200 Library for Spirent Connect SmartBits 200 2000 Installation Guide For new customers please contact Spirent Communications Customer Service abo...

Page 10: ...DLS 400E3 Operating Manual Page 1 6 Spirent Communications 7104000537 03 04 2...

Page 11: ...d to both chassis of the DLS 400E3 and switch the power on 2 Ensure that both of the DLS 400E3 chassis have different IEEE addresses 3 Connect a cable from the computer to chassis 1 If using IEEE 488...

Page 12: ...or even switched off The ATU R and ATU C and their wiring should be physically separated as crosstalk can occur between cabling Configure the cables so that they are not touching and the cable connect...

Page 13: ...B of unit 1connects to side A of unit 2 This one foot RJ 45 cable is provided as an acces sory within the DLS 400E3 package See Section 2 4 4 for RJ 45 connector details 2 4 3 Connecting to Analog De...

Page 14: ...al Note RJ 11 male connectors can also mate to RJ 45 female receptacles These connections are balanced WARNING The maximum input to the DLS 400E3 must not exceed 30 dBm between 50 Hz and 1 kHz Exceedi...

Page 15: ...ased which can inject noise differentially onto the wireline with a 4 KOhm Thevenin output impedance The DLS 400E3 also has 2 50 Ohm BNC external noise inputs located on the rear panel These inputs ar...

Page 16: ...S 400E3 can use any valid IEEE 488 address from 0 to 30 You can change the address by using the DIP switch on the back of the unit The weighting is as follows The following figure shows the default sw...

Page 17: ...tware Requirements You require the following to configure and control a simulator through either the serial port or IEEE 488 interface If using the DLS 400E3 Software provided with the unit a computer...

Page 18: ...le click the icon on your desktop The DLS 400E3 DLS 400E3 Available Configurations window appears Figure 3 1 DLS 400E3 Software Main Window 3 4 Accessing the Control Window s The DLS 400E3 control win...

Page 19: ...ted products connected via RS 232 or the IEEE 488 communication buses optional interface The program automatically displays the connected units in the display area OR 2 Click User Defined then manuall...

Page 20: ...ating Manual Page 3 4 Spirent Communications 7104000537 03 04 2 Figure 3 3 DLS 400E3 Unit 1 Control Window To access the DLS 400E3 chassis 2 controls click the Unit 2 Tab Figure 3 4 DLS 400E3 Unit 2 C...

Page 21: ...e simulator you wish to configure from the drop down list Your choices are 1 30 The default IEEE 488 address is 14 Factory address settings for unit 1 and unit 2 are 14 and 15 respectively OR ComPort...

Page 22: ...or the modems Relay Refresh Check this check box to enable the relay refresh and inject whetting current on the simulated loop Clear this check box to disable the relay refresh feature 3 5 4 Compensat...

Page 23: ...ensate the standard slot settings for the selected test loop 3 5 5 Selecting the Test Loop Note In the Loop Selection area the field names change based on the value you select in the Standard field Fo...

Page 24: ...or each unit Gauge Displays the gauge of wireline for each of the 24 wireline card slots for each unit Tap Check this check box to set the card in the associated slot to represent a bridged tap Clear...

Page 25: ...following graph from the ITU T G shdsl standard illustrates the specified micro interruption Figure 3 5 Micro Interruption Test Circuit The standard specifies two configurations based on the powering...

Page 26: ...ld or use the up and down arrows on the spin box The range is 3 to 360 seconds Test Interval TI Enter the test interval duration in seconds directly in the field or use the up and down arrows on the s...

Page 27: ...ed data Both interfaces use the same command set and pro duce the same results Section 4 2 IEEE 488 Interface and Section 4 3 RS 232 Serial Interface describe features specific to one particular inter...

Page 28: ...signal or both Messages from the system are always terminated with a Line Feed character and the IEEE 488 1 EOI signal Note that some software languages may automatically append a carriage return and...

Page 29: ...haracter 1 stop bit and RTS CTS hardware flow control The RS 232 standard is equivalent to the European V 24 V 28 standards In this manual we use the term RS 232 to refer to both of these two standard...

Page 30: ...he line feed character decimal 10 hex 0A LF To ensure that no characters are left in the receive buffer of the system from a pre vious incomplete command you can send the line feed character by itself...

Page 31: ...a unit i e dB pps mv etc is appended to a number that unit must be valid and not abbreviated Note that the period separates the decimal part of a number 4 5 Command Syntax The host system adheres to t...

Page 32: ...begin with a colon the simulator assumes that the command is at the same level as the previous command For example to set the impulse it is NOT necessary to specify sourceB IMPULSE for each parameter...

Page 33: ...unction Sets the Event Status Enable Register ESER using an integer value from 0 to 255 repre senting a sum of the bits in the following bit map Bits 7 to 0 have values of 128 64 32 16 8 4 2 and 1 res...

Page 34: ...roceed as normal after this command but be prepared to receive the 1 at any time PSC NRf Power on Status Clear Type Status and event command Function Indicates if the unit should clear the Service Req...

Page 35: ...Service Request Enable Query Type Status command Function An integer value representing the value of the Service Request Enable Register is placed in the output queue The possible values are listed in...

Page 36: ...ion with the host system is frozen until all pending operations are completed 4 8 Status Reporting There are two registers that record and report the system status the Status Byte Register STB and the...

Page 37: ...current operation is complete bit 1 Request Control The host system does not have the ability to control the IEEE bus and so this bit is always 0 bit 2 Query Error There was an attempt to read an empt...

Page 38: ...st system can determine when the operation is completed by waiting for SRQ or by reading the status byte with the serial poll or with STB if corre sponding bits are enabled If the program uses the OPC...

Page 39: ...csv files 1 Standards File associates an ID to every possible loop configuration supported by the DLS 400E3 2a DLS 400E3 Standard Slot Settings File associates each ID to generic simulator compensati...

Page 40: ...files associate this ID to exact length settings for each or the 24 slots in unit 1 and unit 2 See the example for ID 19 as shown in Figure 5 2 Figure 5 2 E3 Standard Slot Settings File These setting...

Page 41: ...E3 may be terminated with either a Line Feed character ASCII LF decimal 10 hex 0A an IEEE 488 EOI signal or both RS 232 messages must be terminated with a line feed character Messages from the DLS 400...

Page 42: ...esolution of the parameter The DLS 400E3 system is a slot based configuration and therefore device dependant commands should fol low the structure listed below For automating scripts you can open the...

Page 43: ...18 500 M INLINE SET SLOT 23 500 M INLINE The listed slot settings below are default values and can be varied between inline and tap However Unit 2 Slot 1 can only be set as a bridged tap and cannot be...

Page 44: ...s previous settings You may want to create a loop based command which sets all slots to 0 length prior to setting a new loop Unit 2 SETTINGS SET CHANNEL BYPASS NO SETTING CHANNEL DIRECTION FORWARD SE...

Page 45: ...s allowing a minimum of 15 meter increments A Bypass Card type is returned for any other slot not occupied by a wireline The type Empty Slot is returned if no card wireline or bypass is installed in a...

Page 46: ...n To set a specific slot do the following 1 transmit SET SLOT 1 50 METERS INLINE 2 check that the REMOTE LED is green To send and receive messages with error checking follow these steps 1 set all rele...

Page 47: ...pecified test via SCPI command the following proce dure has to be followed The user should establish the desire Test Interval TI and Signal Period T Then divide TI by T to obtain the number of require...

Page 48: ...ot 1 100 tap Step 2 OPC Step 3 wait SRQ raised or wait for some time Step 4 ESR Step 5 check till the bit 0 of ESR is set Step 6 setting channel interrupt side A 3 Using OPC OPC is Operation Complete...

Page 49: ...ion pin 1 and 2 which resets only the interface when IFC is received The only tool required to change the shunt s position is a screwdriver Ensure that the cover card retainer and card stabilizer hard...

Page 50: ...ption 6 Re install the controller and ensure a tight fit with the connectors on the motherboard 7 At this time plug the unit in and turn it on Run the software that came with the unit and ensure the p...

Page 51: ...rive Suite P S La Mesa CA 91942 Phone 619 697 8790 Fax 619 697 5955 ITU T ADSL2 European Asymmetric Digital Subscriber Line ADSL Transceivers International Telecommunication Union Place des Nations CH...

Page 52: ...DLS 400E3 Operating Manual Page 7 2 Spirent Communications 7104000537 03 04 2...

Page 53: ...gin htm 8 2 Protecting Your Investment Spirent Communications is committed to providing the highest quality products and customer support possi ble An annual calibration is required to ensure that you...

Page 54: ...tandards and Technol ogy N I S T traceable calibration that expires one year from the original ship date With ISO 9000 and other manufacturer specific metrology requirements timely calibrations become...

Page 55: ...e not covered under this warranty the regular repair charges shall apply to any repairs made to the equipment For international customers please contact your local Spirent Communications sales represe...

Page 56: ...DLS 400E3 Operating Manual Page 9 2 Spirent Communications 7104000537 03 04 2...

Page 57: ...e of the unit It is highly recommended that all shipments are marked with labels indicating that the contents are fragile If sending a unit back to the factory ensure that the Return Material Authoriz...

Page 58: ...DLS 400E3 Operating Manual Page 10 2 Spirent Communications 7104000537 03 04 2...

Page 59: ...op 7 MAE 1 0 dB over band used by DUT to a maximum of 90dB in ideal conditions Average Noise Floor 150dBm Hz within the ADSL Band Construction Main chassis plus plug in controller wireline modules and...

Page 60: ...consumption 120 VA max Line Fuses Type T 2A 250V SLOW BLOW 2 required 5 mm x 20 mm Environmental Operating Temperature 10 C to 40 C 50 F to 104 F Storage Temperature 10 C to 40 C 50 F to 104 F Humidi...

Page 61: ...apter 11 SPECIFICATIONS of this manual WARNING To avoid electrical shock do not operate the equipment if it shows any sign of damage to any portion of its exterior surface such as the outer casting or...

Page 62: ...ny devices or cords connected to it are properly grounded 12 2 2 Operating the Unit Do not operate the equipment when its covers or panels have been removed Do not operate equipment if an interruption...

Page 63: ...ications Page 12 3 7104000537 03 04 2 12 3 Symbols When any of these symbols appear on the unit this is their meaning EQUIPOTENTIALITY FUNCTIONAL EARTH TERMINAL PROTECTIVE GROUNDING CONDUCTOR TERMINAL...

Page 64: ...DLS 400E3 Operating Manual Page 12 4 Spirent Communications 7104000537 03 04 2...

Reviews: