4-272
Flow-X─ Function Reference
Flow-X General functions - fxSarasota_M
Where:
ρ
C
The corrected density
kg/m3
d
0
Obtained from the calibration certificate
kg/m3
0
Obtained from the calibration certificate
s
K
Obtained from the calibration certificate
-
d
0
Obtained from the calibration certificate
-
p
COE
F
Obtained from the calibration certificate
s/bar
t
COE
F
Obtained from the calibration certificate
s/
°
C
t
Line temperature
°C
t
CAL
Reference temperature
°C
p
Line temperature
bar(g)
p
CAL
Reference pressure
bar(g)
C
Time periodic input corrected for temperature and
pressure
s
τ
The time period in µS
s