background image

Sponsler Co., Inc.

SP717 Modulated Carrier Amplifier

pg 3

DOC#” MN-717-A

BENCH TEST CALIBRATION PROCEDURE

Required Equipment:

 Power Supply 6-28 VDC, Frequency Generator, Frequency Counter,

               Oscilloscope

Test Procedure:

A)

Connect RF pickup coil to J1-1,2

B)

Connect power supply positive & negative leads to J1-3,5 respectively

C)

Connect oscilloscope  positive & negative leads to J1-1,2 respectively

D)

Install jumper @ JU1

E)

Turn power supply “ON”, observe LED D1 illuminates and oscilloscope
displays a 50khz + 5Khz carrier sinewave

F)

Observe carrier amplitude of 6Vp-p nominal

G)

Connect oscilloscope positive & negative leads to J1-4,5 respectively

H)

Connect frequency generator positive & negative leads to J1-1,2 respectively.
Set function to squarewave, amplitude to 5Vp-p & frequency to 1-3500 Hz.

 I)

LED D2 illuminates & the oscilloscope displays a squarewave whose
frequency is that of step H with a positive amplitude .6V < power supply
positive potential and a negative amplitude .6V > power supply negative
potential.

J)

Observing oscilloscope display momentarily short pins J1-4 & 5 together. The
oscilloscope will display a squarewave with an amplitude of .6V or less.
When the short is removed the oscilloscope will display a normal amplitude
squarewave.

Reviews: