8
RUNNER srl
Via G. di Vittorio n. 391 - 41032 Cavezzo (MO) Italia Tel. +39 0535 58447
1.1.
Symbols
Conformity to European
Community Rules
Electrical grounding
Producer
Warning of foot squashing
Date of production
Electrical grounding
Attention, please
read the manual
connecting point
RAEE waste
Warning of electrical voltage
B Type machine
Catalogue code
Read carefully the manual
The Directive tells as "applied parts" the components of a medical device that in the normal use are directly in
contact with the human body. They can be made up of electrodes, sensors applied to the patient, catheters
containing physiological fluids or simply from casing of the device. The contact of the applied part with the
patient obviously makes greater risk compared to a part of the device that the patient may be able to touch,
voluntarily or unintentionally, directly or indirectly, but however occasionally. From every device, despite the
high impedance insulation, flows a small current that disperses towards the ground, on the casing and in the
patient. The current of dispersion to the ground is thatfrom the device connected to power supply, through
or through along the insulation, walks the protective conductor to the ground. The scatter current on the
casing is the current that travels the casing of a device in normal use (excluding applied parts) accessible to
the patient or operator, crossing the subject connection between the casing and the ground or with two points
of the casing. Applied part can be defined as the set of all device parts that in normal use is essential to put in
contact with the patient so that the device can perform its function or that they can come into contact with
the patient or that they need to be necessarily touched by the patient. Depending on the currents dispersion
in normal conditions and first fault to the ground, on the casing and on the patient, our electromedical devices
are classified as type B applicances.
Summary of Contents for Runner RUN2011/T-PC
Page 1: ...SPORT TEC LAUFBAND RUN MED ANLEITUNG Art Nr 88812 Kategorie Laufb nder...
Page 2: ......
Page 3: ......
Page 93: ...Cod 0103...
Page 94: ......
Page 95: ......
Page 183: ......
Page 268: ......
Page 269: ...Cod 0100...