Model SR755KG Series Wheelchair Scale
Operating and Service Manual - S/N 1913+
Part No.: MAN755KG_100929
Page 9 of 16
Model SR755KG Series Wheelchair Scale
Operating and Service Manual - S/N 1913+
Part No.: MAN755KG_100929
Page 9 of 16
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
sri@srinstruments.com
Copyright 2010
S
Instruments, Inc.
email:
sri@srinstruments.com
Copyright 2010
S
Instruments, Inc.
BASIC SYSTEM OPERATION cont’d
STEP 1
: Place the empty wheelchair (with any blankets or pillows that the patient may have
with them) on the scale platform. Then press the zero button.
NOTE
:
The display will shut off after 35 seconds. The zero (TARE) will still be stored
in memory.
STEP 2
: Remove the wheelchair from the platform. Place the patient in the wheelchair.
STEP 3
: Wheel the patient with wheelchair onto the platform. Lock the patient’s wheelchair
brake to prevent movement. Press the weigh button. The display will show the patient’s weight
only.
CONTINUOUS WEIGH
In this default mode, the weighing surface remains active. Press the “
HOLD
” button once to lock
the displayed reading and store it in memory as the “last weight” for recall later if needed.
AUTO-HOLD
This mode is for patients unable to remain still for the weighing procedure. It locks, stores, and
displays the patient’s weight as soon as the “
WEIGH
” button is pressed once.
Note
: No weight
will be displayed until the button is pressed.
To enable this mode, BEFORE zeroing the system, press and hold the “
HOLD
” button for
approximately five (5) seconds until the display reads “
AUTO-HOLD ENABLED
”.
To return to CONTINUOUS WEIGH mode when finished, press and hold the “
HOLD
” button for
approximately five (5) seconds until the display reads “
CONTINUOUS WEIGH
”.
BATTERY REPLACEMENT
STEP 1
: The display will read “
REPLACE BATTERY
”.
STEP 2
: (Figure 5) Place the scale on its side and open the Battery Compartment
cover located
in the bottom of the scale between the wheels.
STEP 3
: Remove and replace ALL six
(6) “D” cell batteries. Refer to diagram on
the Battery Compartment cover for
placement.
STEP 4
: Press
the
“
WEIGH
” button to
confirm display is working.
STEP 5
: Tightly close the cover.
STEP 6
: Zero the system.
Figure 5: Battery Compartment
CAUTION