Model SR7010
i
Wall Mount Platform Scale
Operating and Service Manual
Part No. MAN7010
i
_150908
Page 10 of 14
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
sri@srinstruments.com
Copyright 2015
S
Instruments, Inc.
L O W
L IM IT
A P P L IE D
L O A D
H IG H
L IM IT
9 9 .9
1 0 0 .0
1 0 0 .1
1 9 9 .8
2 0 0 .0
2 0 0 .2
2 9 9 .7
3 0 0 .0
3 0 0 .3
3 9 9 .6
4 0 0 .0
4 0 0 .4
4 9 9 .5
5 0 0 .0
5 0 0 .5
5 9 9 .4
6 0 0 .0
6 0 0 .6
6 9 9 .3
7 0 0 .0
7 0 0 .7
7 9 9 .2
8 0 0 .0
8 0 0 .8
8 9 9 .1
9 0 0 .0
9 0 0 .7
9 9 9 .0
1 0 0 0 .0
1 0 0 1 .0
C A L IB R A T IO N
T O L E R A N C E T A B L E
CALIBRATION
NOTE:
Ensure that nothing is in contact with the scale system during
this procedure. Remove hands from the system when noting the
displayed calibration results.
CHECKING CALIBRATION
STEP 1:
Select known calibrated weights, traceable to NIST.
NOTE: DO NOT USE
barbells or uncalibrated weights.
STEP 2:
Zero the scale by pressing and holding the left
ZERO
button.
STEP 3:
Place the calibrated weight on the scale. Wait for scale to
stabilize; note scale reading.
STEP 4:
Scale readings should be within Calibration Tolerance Table
(
Figure 3
)
Figure 3: Calibration Table
Figure 4: Calibration Switch Diagram
IMPORTANT
CALIBRATION Qualified service personnel only should perform this procedure. The
SR7010
i
load cells have no user serviceable components and should not be tampered with
for any reason. Re-calibration is generally not required, but should be verified
periodically to ensure accuracy. The recommendation for calibration check is at least
once every 12 months, or as individual maintenance policy requires.
CALIBRATION SWITCH
CAUTION
The integrated circuits and semiconductors on
the printed circuit boards may be damaged
by electrostatic discharge (ESD). Be sure to
use proper handling precautions at all times.
Continued next page