background image

Model SRV930KG Series Medium Capacity Counter Top Scale 
Operating and Service Manual  
Part No.:  MANSRV930KG_201102 

 

Page 7 of 14 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

email:  

sri@srinstruments.com

        

Copyright 

2020

 

S

Instruments, Inc. 

TARE/ON/OFF 

Kg/Lb 

MEMORY 

BUTTON FUNCTIONS  

 

 

 

 

The “

TARE/ON/OFF

” button is used to turn the scale on, off, to tare the 

weigh, or to zero the display.   

 

Holding the “

TARE/ON/OFF

” button for two seconds will manually shut 

down the scale.   

 

 
Button is disabled on this model  
 
 
The  “

MEMORY

”  button  is  used  to  recall  the  weight  of  the  patient  in 

memory.   

 
 

 

Figure 2:  Display Button Functions 

Summary of Contents for SRV930KG Series

Page 1: ...RV930KG_201102 Page 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by...

Page 2: ...ABLE OF CONTENTS TABLE OF FIGURES 2 PACKING CHECKLIST MODEL SRV930KG 3 ASSEMBLY 3 ACCESSORIES 3 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE AND TRANSPORTATION 4 CLEANING AND DISINFECTING 5 SPECIFICA...

Page 3: ...heck parts against the PACKING CHECKLIST If there are any missing or damaged parts please call the Service Hotline 1 800 654 6360 STEP 2 Figure 1 Turn scale over Slide open Battery Compartment Cover I...

Page 4: ...AC adaptor SRV930AC The animal s weight is displayed on a 4 digit LCD display INTENDED USE The SRV930KG is designed for use as an animal weighing system for animals not exceeding 20 kilograms Please...

Page 5: ...contact areas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially avail...

Page 6: ...SOLUTION 0 01 kg AUTO POWER DOWN Adjustable to 120 180 240 300 seconds and OFF MEMORY While active freezes display data and stores it in memory LAST WEIGHT RECALL MEMORY button recalls last stored wei...

Page 7: ...es com email sri srinstruments com Copyright 2020 SInstruments Inc TARE ON OFF Kg Lb MEMORY BUTTON FUNCTIONS The TARE ON OFF button is used to turn the scale on off to tare the weigh or to zero the di...

Page 8: ...to subtract any additional items example receptacle from the patient s Net weight MEMORY Press the MEMORY button to store the last weight for recall if needed To view the last stored weight in memory...

Page 9: ...til the word END is displayed Press the MEMORY button to save your changes the display will return to normal mode BATTERY REPLACEMENT The low battery indicator will display on LCD display when battery...

Page 10: ...rams mode STEP 4 Place 20 kg dead weight on the scale platform Display will show a span count between 2000 16000 STEP 5 Press TARE ON OFF button one 1 time Display will read CAL and then read 20 00 kg...

Page 11: ...the total weight exceeds the maximum capacity of the scale removed weight from the surface Memory will not Zero Memory will hold the last stored weight until it is replaced with another stored weight...

Page 12: ...tions or liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said arti...

Page 13: ...Scale Operating and Service Manual Part No MANSRV930KG_201102 Page 13 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinst...

Page 14: ...Manual Part No MANSRV930KG_201102 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 S...

Reviews: