32
2 Operation
The
Pending
indicator will light, indicating that a press of the
button is required to complete the
Clear ERR
operation. All errors will
be cleared, and
ERR
will turn off. The Status Byte is not cleared.
Any button other than
will cancel the
Clear ERR
operation.
2.2.6
Phase
The
PHASE
block of the front panel consists only of the
Rel
Phase
button.
For discussion of the
Phase
setting, see Section 2.2.2.2.
The
Rel
Phase
button is used to set the phase = 0
°
reference from which the
phase setpoint is measured. When relative phase is inactive, the
Rel
LED is unlit and the displayed
Phase
setting represents the commanded
phase difference between the source signal and the selected
Control
track, as measured by the shaft encoder.
When
Rel
Phase
is pressed, the currently displayed
Phase
setting is zeroed
(similar to pressing the tare button on a scale) and the
Rel
indicator
illuminates. All subsequent changes to the phase are relative to this new
zero.
Pressing
Rel
Phase
again returns the displayed phase setting to an absolute
measurement. See the example provided with the
command.
2.2.7
Motor
The
Run /
Stop
button toggles power to the chopper head motor. A long press
is required to start the motor, while a short press will bring it to a stop.
At startup, the rotor shaft orientation must first be indexed, and so you
may notice several low speed rotations in alternate directions until the
index position is found. Following indexing, the motor always spins
up to a speed of approximately 5 rev/s in order to survey the installed
blade (count the slots of the inner and outer tracks), even if the setpoint
frequency requires a speed less than 5 rev/s (including the 0 Hz
shutter
mode
). Once the blade has been surveyed, the motor will spin up to the
necessary speed to achieve frequency and phase lock. Typically a few
seconds are required to achieve phase lock (up to 15 s, depending on the
frequency).
When
Run /
Stop
is pressed to stop the motor, braking current will be applied
to bring the chopper wheel to a smooth but rapid halt. The
Run
LED
will only shut off after the chopper head has been de‑energized, so there
will be some delay between the “Stop” button press and the
Run
and
Stop
LED indicator update.
SR542 Precision Optical Chopper